210 Series Transmitter with External Electrochemical Sensor INSTRUCTIONS Installation and Operation of the AMC-210 Series Transmitter with Exter
AMC-210 Transmitter with external electrochemical sensor 8 3.2 CABLE SELECTION AND WIRING Connection should be made using 2-conductor, shielded ca
AMC-210 Transmitter with external electrochemical sensor 9 FIGURE 3: Cable selection graph
AMC-210 Transmitter with external electrochemical sensor 10 3.2.1 TRANSMITTER TO MONITOR WIRING The transmitter output (–, +) terminals connect to
AMC-210 Transmitter with external electrochemical sensor 11 3.2.2 INTERFACING TO COMPUTER, DATALOGGER, OR NON-AMC MONITOR All Armstrong sensor/tra
AMC-210 Transmitter with external electrochemical sensor 12 4 OPERATION AND CALIBRATION 4.1 OPERATION The AMC-210 series sensor/transmitter is fac
AMC-210 Transmitter with external electrochemical sensor 13 5. Extended warranty calibration program For all above options, please contact AMC fo
AMC-210 Transmitter with external electrochemical sensor 14 RECALIBRATION For full recalibration adjustments, follow the Set-Up procedure steps 1 t
AMC-210 Transmitter with external electrochemical sensor 15 5 PREVENTIVE MAINTENANCE 5.1 GENERAL The sensor/transmitter unit should be brushed or
AMC-210 Transmitter with external electrochemical sensor 16 FIGURE 7: Sensor replacement and wiring procedure.
AMC-210 Transmitter with external electrochemical sensor 17 6 ADDENDUM (FOR OXYGEN ONLY) This AMC-210 Series (2-wire sensor/transmitter with exter
NOTE This page intentionally left blank.
AMC-210 Transmitter with external electrochemical sensor 1 TABLE OF CONTENTS
AMC-210 Transmitter with external electrochemical sensor 2 NOTE This page intentionally left blank.
AMC-210 Transmitter with external electrochemical sensor 3 1 WARRANTY The AMC-210 transmitter is warranted against defects in material and workman
AMC-210 Transmitter with external electrochemical sensor 4 2 PRODUCT INFORMATION The AMC-210 series sensor/transmitter is designed to provide cont
AMC-210 Transmitter with external electrochemical sensor 5 2.2 FACTORY CALIBRATION Gas Type……………………………………………….…..…. Zero Gas, at 4 mA Signal…………
AMC-210 Transmitter with external electrochemical sensor 6 2.3 HOUSING OPTIONS The AMC-210 series sensor/transmitter units are available in the hou
AMC-210 Transmitter with external electrochemical sensor 7 3 INSTALLATION 3.1 LOCATION AND MOUNTING Mount the sensor/transmitter on a solid, non-v